The figure
shown on the right is an autocorrelation trace from one of our
standard CPA-2001. The autocorrelation trace was collected using
our model AC-150 autocorrelator and associated ODLDA data acqusition
software. We took advantage of the long-range capability of the
AC-150 to search for the presence of secondary pulses around the
main pulse.
Secondary pulses are often found on poorly aligned femtosecond
systems. Secondary pulses that are located far away from the main
pulse can be found using the traditional photodetector and oscilloscope.
Secondary pulses located too close to the primary pulse for differentiation
using traditional photodetector and oscilloscope can be challenging
to detect. The user can be induced into a false sense of security
(but we bet your experiment will be affected by these pesty secondary
pulses!).
Using the AC-150 we
were able to confirm the absence of secondary pulses over a range
exceeding 100 ps. See
Specifications
Our best measurement of real pulsewidth
to date!* |

Warranty
Oscillator parts, including
the diode laser, are warranted for five (5) years. For details,
contact our Sales Department (sales@cmxr.com)
*Not
representative of actual end user results
|
|